<a id="datasheet-js220plus"></a>

# Joulescope JS220+ Datasheet

The Joulescope™ JS220+ is a limited-edition, performance-enhanced version of
the Joulescope JS220.  The JS220+ achieves ±1 nanoamp accuracy with
exceptional temperature stability, all while providing the same best-in-class
high-dynamic current range measurement as the standard JS220.

The [Joulescope JS220+ User’s Guide](https://docs.joulescope.com/en/products/js220plus/guide/index.md) describes the
JS220+ improvements, and the
[Joulescope JS220 User’s Guide](https://docs.joulescope.com/en/products/js220/guide/index.md) describes how to
use the instrument.  The specification values are also available as
[js220plus_datasheet.yaml](https://docs.joulescope.com/en/products/js220plus/js220plus_datasheet.yaml),
which holds the differences from
[js220_datasheet.yaml](https://docs.joulescope.com/en/products/js220/js220_datasheet.yaml).

<!--
Include-only fragment: the JS220+ specifications.  Included by
guide/specifications.md and datasheet.md.  No labels here, since
each label would be defined twice.  Table values: js220plus.yaml.
-->

The JS220+ specifications are the same as the JS220, except for the current
measurement, voltage measurement and parasitic effects specifications.
Resolution is given for 10 kHz measurement bandwidth (20 ksps).  The JS220+
includes a safety banana jack front panel rather than binding posts.

## Absolute maximum ratings

|                                                                                      | Min       | Max       | Unit   |
|--------------------------------------------------------------------------------------|-----------|-----------|--------|
| USB voltage                                                                          | -0.3      | 6         | V      |
| Max voltage applied between any port (Current +, Current -, Voltage +, Voltage -)    | -20       | 20        | V      |
| Max voltage to USB ground from any port (Current +, Current -, Voltage +, Voltage -) | -48       | 48        | V      |
| Input current (powered, burst)                                                       | -10       | 10        | A      |
| Input current (powered, sustained)                                                   | -3        | 3         | A      |
| Input current (unpowered)                                                            | N/A (off) | N/A (off) | A      |
| Storage temperature                                                                  | -40       | 80        | °C     |
| Operating temperature                                                                | 0         | 50        | °C     |
| Relative humidity (non-condensing)                                                   | 0         | 95        | %      |
| Altitude from sea level                                                              | -100      | 2500      | m      |
| Max voltage applied to GPI relative to USB ground                                    | -10       | 10        | V      |

## Mechanical

### Instrument only

|                                                    |   Typical | Unit   |
|----------------------------------------------------|-----------|--------|
| Weight                                             |       190 | grams  |
| Width                                              |        99 | mm     |
| Length (including binding posts and BNC connector) |       126 | mm     |
| Height                                             |        28 | mm     |

### Complete product kit

|        |   Typical | Unit   |
|--------|-----------|--------|
| Weight |       510 | grams  |
| Width  |       190 | mm     |
| Length |       200 | mm     |
| Height |        60 | mm     |

## Recommended operating conditions

|                                                | Test condition                   | Min             |   Typical | Max            | Unit   |
|------------------------------------------------|----------------------------------|-----------------|-----------|----------------|--------|
| Operating temperature                          |                                  | 10              |           | 40             | °C     |
| USB supply voltage                             |                                  | 4.5             |         5 | 5.5            | V      |
| USB supply current                             |                                  |                 |           | 500            | mA     |
| Voltage applied across Voltage + and Voltage - |                                  | -14.5           |           | 14.5           | V      |
| Voltage applied across Current + and Current - | Current range off                | -15             |           | 15             | V      |
| Input current                                  | continuous                       | -3              |           | 3              | A      |
| Input current pulse                            | < 50 milliseconds, < 3A average  | -10<sub>1</sub> |           | 10<sub>1</sub> | A      |
| Input current pulse                            | < 250 milliseconds, < 3A average | -5<sub>1</sub>  |           | 5<sub>1</sub>  | A      |
1. Suitable for handling inrush currents when connecting capacitive loads.

## Specification conditions

| Temperature          | 23 °C ± 10 °C                 |
|----------------------|-------------------------------|
| Humidity             | 20% to 80% RH, non-condensing |
| Calibration interval | 1 year<sub>1</sub>            |
1. Recommended to comply with industry-standard practices, but the Joulescope JS220 is designed not to need recalibration.

## Current measurement specifications

| Range                    | Accuracy                | Resolution   | Impedance   | Guaranteed   |
|--------------------------|-------------------------|--------------|-------------|--------------|
| 10 A (> 2 A)<sub>1</sub> | ±0.25% ±1.5 mA          | 350 µA       | 0.01 Ω      | ±9.7 A       |
| 10 A (≤ 2 A)             | ±0.1% ±300 µA           | 75 µA        | 0.01 Ω      | ±2.05 A      |
| 180 mA                   | ±0.1% ±30 µA            | 6.5 µA       | 0.11 Ω      | ±185 mA      |
| 18 mA                    | ±0.1% ±3.0 µA           | 650 nA       | 1.11 Ω      | ±18.5 mA     |
| 1.8 mA                   | ±0.1% ±300 nA           | 65 nA        | 11.1 Ω      | ±1.85 mA     |
| 180 µA                   | ±0.1% ±30 nA            | 6.5 nA       | 111 Ω       | ±185 µA      |
| 18 µA                    | ±0.1% ±3 nA<sub>2</sub> | 0.65 nA      | 1111 Ω      | ±18.5 µA     |
1. Max 3 A sustained.  See the recommended operating conditions for higher current durations.
2. ±1.0 nA for ±5 °C over 24 hours with trim offset calibration.

Test current measurement bandwidth by varying the effective load impedance.
Due to load capacitance, altering the source voltage is not a valid method of
testing current bandwidth.

The JS220 is sensitive to RF radiation which induces real current in test
setups.  When tested with EN61326-1 and 3 V RMS/m, the JS220 measured up to
10 µA (observed, not guaranteed) of current at several frequencies.  The
exact frequency susceptibility varies with cable length and configuration.

## Voltage measurement specifications

| Range   | Accuracy      | Resolution   | Guaranteed   |
|---------|---------------|--------------|--------------|
| 15 V    | ±0.1% ±1 mV   | 500 µV       | ±14.5 V      |
| 2 V     | ±0.1% ±150 µV | 75 µV        | ±2.2 V       |

The JS220 is sensitive to RF radiation which induces real voltages in test
setups.  When tested with EN61326-1 and 3 V RMS/m, the JS220 error increased
by 10 mV (observed, not guaranteed) at several frequencies.  The exact
frequency susceptibility varies with cable length and configuration.

## General specifications

|                        | Test condition   |   Min |   Typical |   Max | Unit   |
|------------------------|------------------|-------|-----------|-------|--------|
| Sampling frequency     | 25 °C            |       |   2000000 |       | sps    |
| Sampling frequency     | 25 °C            |   -25 |           |    25 | ppm    |
| Measurement bandwidth  | -3 dB            |   280 |       310 |   350 | kHz    |
| Front panel insertions |                  |       |           |   100 |        |

## Parasitic effects specifications

|                                                  | Test condition                 |   Min | Typical   | Max   | Unit   |
|--------------------------------------------------|--------------------------------|-------|-----------|-------|--------|
| Current leakage (Voltage + to -)                 | 25 °C, ±3.3V                   |       |           | 200   | pA     |
| Current leakage (Voltage + to -)                 | 25 °C, ±5V                     |       |           | 200   | pA     |
| Current leakage (Voltage + to -)                 | 25 °C, ±14V                    |       |           | 300   | pA     |
| Shunt resistor voltage, max                      | -2 to 2 A, static              |       | 20        | 21    | mV     |
| Burden voltage (Current + to -)                  | 1 A, banana jacks              |     0 | 25        | 35    | mV     |
| Burden voltage (Current + to -)                  | Shunt resistor, < ±2 A         |       |           | ±20   | mV     |
| Shunt switching time                             | Over-range, < 200 ns rise time |     0 | 1.0       | 1.5   | µs     |
| Measurement gap following a current range switch |                                |       |           | 0     | µs     |
| Input bias current Voltage+ & -                  | 25 °C                          |       | ±5        | ±20   | pA     |
| Input bias current, Current + & -                | 25 °C, 18 µA current range     |       | ±60       | ±600  | pA     |

## Software-controlled fuse specifications

|                                  | Test condition   | Min   | Typical   | Max   | Unit   |
|----------------------------------|------------------|-------|-----------|-------|--------|
| Current leakage (Current + to -) | 25 °C, 3.3V      |       | ±5        |       | nA     |
| Current leakage (Current + to -) | 25 °C, 5V        |       | ±7        |       | nA     |
| Current leakage (Current + to -) | 25 °C, 15V       |       | ±28       | ±5000 | nA     |
| Max fuse engage                  | 5 A              |       | 0.293     |       | s      |
| Max fuse engage                  | 10 A             |       | 0.050     |       | s      |

## General purpose inputs and outputs

|                           | Test condition        | Min        | Typical   | Max        | Unit   |
|---------------------------|-----------------------|------------|-----------|------------|--------|
| Internal 3.3V reference   |                       | 3.135      | 3.3       | 3.465      | V      |
| External reference        | Vref < (VUSB - 0.5 V) | Vref - 0.1 | Vref      | Vref + 0.1 | V      |
| Input leakage per pin     | 25 °C                 |            |           | ±10        | nA     |
| Input leakage per pin     |                       |            |           | ±100       | nA     |
| Output voltage high       | -100 µA               | Vcco - 0.1 |           |            | V      |
| Output impedance          | 10 mA                 | 45         | 50        | 80         | Ω      |
| GPI sample rate           |                       |            | 2         |            | MSPS   |
| Continuous output current | Per pin               |            |           | ±50        | mA     |
| Total power dissipation   | Over output impedance |            |           | 0.25       | W      |

The Joulescope uses the TI SN74LVC2T45 for both the GPIs and GPOs.  See the
[SN74LVC2T45 datasheet](http://www.ti.com/lit/gpn/sn74lvc2t45) for more
information.  Both the GPIs and GPOs are protected with the TI TPD4E1U06 for
ESD protection.