# Specifications

This section only gives the specification changes relative to the
[Joulescope JS220 User’s Guide](https://docs.joulescope.com/en/products/js220/guide/index.md).  All other
specifications remain the same as the JS220.  The
[JS220+ datasheet](https://docs.joulescope.com/en/products/js220plus/datasheet.md#datasheet-js220plus) lists the complete
specifications.

## Current measurement specifications

This table replaces the JS220
[current measurement specifications](https://docs.joulescope.com/en/products/js220/guide/specifications.md#js220-current-measurement-specifications).

| Range                    | Accuracy                | Resolution   | Impedance   | Guaranteed   |
|--------------------------|-------------------------|--------------|-------------|--------------|
| 10 A (> 2 A)<sub>1</sub> | ±0.25% ±1.5 mA          | 350 µA       | 0.01 Ω      | ±9.7 A       |
| 10 A (≤ 2 A)             | ±0.1% ±300 µA           | 75 µA        | 0.01 Ω      | ±2.05 A      |
| 180 mA                   | ±0.1% ±30 µA            | 6.5 µA       | 0.11 Ω      | ±185 mA      |
| 18 mA                    | ±0.1% ±3.0 µA           | 650 nA       | 1.11 Ω      | ±18.5 mA     |
| 1.8 mA                   | ±0.1% ±300 nA           | 65 nA        | 11.1 Ω      | ±1.85 mA     |
| 180 µA                   | ±0.1% ±30 nA            | 6.5 nA       | 111 Ω       | ±185 µA      |
| 18 µA                    | ±0.1% ±3 nA<sub>2</sub> | 0.65 nA      | 1111 Ω      | ±18.5 µA     |
1. Max 3 A sustained.  See the recommended operating conditions for higher current durations.
2. ±1.0 nA for ±5 °C over 24 hours with trim offset calibration.

## Voltage measurement specifications

This table replaces the JS220
[voltage measurement specifications](https://docs.joulescope.com/en/products/js220/guide/specifications.md#js220-voltage-measurement-specifications).

| Range   | Accuracy      | Resolution   | Guaranteed   |
|---------|---------------|--------------|--------------|
| 15 V    | ±0.1% ±1 mV   | 500 µV       | ±14.5 V      |
| 2 V     | ±0.1% ±150 µV | 75 µV        | ±2.2 V       |

## Parasitic effects specifications

This table replaces the JS220
[parasitic effects specifications](https://docs.joulescope.com/en/products/js220/guide/specifications.md#js220-parasitic-effects-specifications).

|                                                  | Test condition                 |   Min | Typical   | Max   | Unit   |
|--------------------------------------------------|--------------------------------|-------|-----------|-------|--------|
| Current leakage (Voltage + to -)                 | 25 °C, ±3.3V                   |       |           | 200   | pA     |
| Current leakage (Voltage + to -)                 | 25 °C, ±5V                     |       |           | 200   | pA     |
| Current leakage (Voltage + to -)                 | 25 °C, ±14V                    |       |           | 300   | pA     |
| Shunt resistor voltage, max                      | -2 to 2 A, static              |       | 20        | 21    | mV     |
| Burden voltage (Current + to -)                  | 1 A, banana jacks              |     0 | 25        | 35    | mV     |
| Burden voltage (Current + to -)                  | Shunt resistor, < ±2 A         |       |           | ±20   | mV     |
| Shunt switching time                             | Over-range, < 200 ns rise time |     0 | 1.0       | 1.5   | µs     |
| Measurement gap following a current range switch |                                |       |           | 0     | µs     |
| Input bias current Voltage+ & -                  | 25 °C                          |       | ±5        | ±20   | pA     |
| Input bias current, Current + & -                | 25 °C, 18 µA current range     |       | ±60       | ±600  | pA     |

Resolution is given for 10 kHz measurement bandwidth (20 ksps).

The JS220+ includes a safety banana jack front panel rather than binding
posts.