Joulescope JS220+ Datasheet#
The Joulescope™ JS220+ is a limited-edition, performance-enhanced version of the Joulescope JS220. The JS220+ achieves ±1 nanoamp accuracy with exceptional temperature stability, all while providing the same best-in-class high-dynamic current range measurement as the standard JS220.
The Joulescope JS220+ User’s Guide describes the JS220+ improvements, and the Joulescope JS220 User’s Guide describes how to use the instrument. The specification values are also available as js220plus.yaml, which holds the differences from js220.yaml.
The JS220+ specifications are the same as the JS220, except for the current measurement, voltage measurement and parasitic effects specifications. Resolution is given for 10 kHz measurement bandwidth (20 ksps). The JS220+ includes a safety banana jack front panel rather than binding posts.
Absolute maximum ratings#
Min |
Max |
Unit |
|
|---|---|---|---|
USB voltage |
-0.3 |
6 |
V |
Max voltage applied between any port (Current +, Current -, Voltage +, Voltage -) |
-20 |
20 |
V |
Max voltage to USB ground from any port (Current +, Current -, Voltage +, Voltage -) |
-48 |
48 |
V |
Input current (powered, burst) |
-10 |
10 |
A |
Input current (powered, sustained) |
-3 |
3 |
A |
Input current (unpowered) |
N/A (off) |
N/A (off) |
A |
Storage temperature |
-40 |
80 |
°C |
Operating temperature |
0 |
50 |
°C |
Relative humidity (non-condensing) |
0 |
95 |
% |
Altitude from sea level |
-100 |
2500 |
m |
Max voltage applied to GPI relative to USB ground |
-10 |
10 |
V |
Mechanical#
Instrument only#
Typical |
Unit |
|
|---|---|---|
Weight |
190 |
grams |
Width |
99 |
mm |
Length (including binding posts and BNC connector) |
126 |
mm |
Height |
28 |
mm |
Complete product kit#
Typical |
Unit |
|
|---|---|---|
Weight |
510 |
grams |
Width |
190 |
mm |
Length |
200 |
mm |
Height |
60 |
mm |
Recommended operating conditions#
Test condition |
Min |
Typical |
Max |
Unit |
|
|---|---|---|---|---|---|
Operating temperature |
10 |
40 |
°C |
||
USB supply voltage |
4.5 |
5.0 |
5.5 |
V |
|
USB supply current |
500 |
mA |
|||
Voltage applied across Voltage + and Voltage - |
-14.5 |
14.5 |
V |
||
Voltage applied across Current + and Current - |
Current range off |
-15 |
15 |
V |
|
Input current |
continuous |
-3 |
3 |
A |
|
Input current pulse |
< 50 milliseconds, < 3A average |
-101 |
101 |
A |
|
Input current pulse |
< 250 milliseconds, < 3A average |
-51 |
51 |
A |
Suitable for handling inrush currents when connecting capacitive loads.
Specification conditions#
Temperature |
23 °C ± 10 °C |
Humidity |
20% to 80% RH, non-condensing |
Calibration interval |
1 year1 |
Recommended to comply with industry-standard practices, but the Joulescope JS220 is designed not to need recalibration.
Current measurement specifications#
Range |
Accuracy |
Resolution |
Impedance |
Guaranteed |
|---|---|---|---|---|
10 A (> 2 A)1 |
±0.25% ±1.5 mA |
350 µA |
0.01 Ω |
±9.7 A |
10 A (≤ 2 A) |
±0.1% ±300 µA |
75 µA |
0.01 Ω |
±2.05 A |
180 mA |
±0.1% ±30 µA |
6.5 µA |
0.11 Ω |
±185 mA |
18 mA |
±0.1% ±3.0 µA |
650 nA |
1.11 Ω |
±18.5 mA |
1.8 mA |
±0.1% ±300 nA |
65 nA |
11.1 Ω |
±1.85 mA |
180 µA |
±0.1% ±30 nA |
6.5 nA |
111 Ω |
±185 µA |
18 µA |
±0.1% ±3 nA2 |
0.65 nA |
1111 Ω |
±18.5 µA |
Max 3 A sustained. See the recommended operating conditions for higher current durations.
±1.0 nA for ±5 °C over 24 hours with trim offset calibration.
Test current measurement bandwidth by varying the effective load impedance. Due to load capacitance, altering the source voltage is not a valid method of testing current bandwidth.
The JS220 is sensitive to RF radiation which induces real current in test setups. When tested with EN61326-1 and 3 V RMS/m, the JS220 measured up to 10 µA (observed, not guaranteed) of current at several frequencies. The exact frequency susceptibility varies with cable length and configuration.
Voltage measurement specifications#
Range |
Accuracy |
Resolution |
Guaranteed |
|---|---|---|---|
15 V |
±0.1% ±1 mV |
500 µV |
±14.5 V |
2 V |
±0.1% ±150 µV |
75 µV |
±2.2 V |
The JS220 is sensitive to RF radiation which induces real voltages in test setups. When tested with EN61326-1 and 3 V RMS/m, the JS220 error increased by 10 mV (observed, not guaranteed) at several frequencies. The exact frequency susceptibility varies with cable length and configuration.
General specifications#
Test condition |
Min |
Typical |
Max |
Unit |
|
|---|---|---|---|---|---|
Sampling frequency |
25 °C |
2000000 |
sps |
||
Sampling frequency |
25 °C |
-25 |
25 |
ppm |
|
Measurement bandwidth |
-3 dB |
280 |
310 |
350 |
kHz |
Front panel insertions |
100 |
Parasitic effects specifications#
Test condition |
Min |
Typical |
Max |
Unit |
|
|---|---|---|---|---|---|
Current leakage (Voltage + to -) |
25 °C, ±3.3V |
200 |
pA |
||
Current leakage (Voltage + to -) |
25 °C, ±5V |
200 |
pA |
||
Current leakage (Voltage + to -) |
25 °C, ±14V |
300 |
pA |
||
Shunt resistor voltage, max |
-2 to 2 A, static |
20 |
21 |
mV |
|
Burden voltage (Current + to -) |
1 A, banana jacks |
0 |
25 |
35 |
mV |
Burden voltage (Current + to -) |
Shunt resistor, < ±2 A |
±20 |
mV |
||
Shunt switching time |
Over-range, < 200 ns rise time |
0 |
1.0 |
1.5 |
µs |
Measurement gap following a current range switch |
0 |
µs |
|||
Input bias current Voltage+ & - |
25 °C |
±5 |
±20 |
pA |
|
Input bias current, Current + & - |
25 °C, 18 µA current range |
±60 |
±600 |
pA |
Software-controlled fuse specifications#
Test condition |
Min |
Typical |
Max |
Unit |
|
|---|---|---|---|---|---|
Current leakage (Current + to -) |
25 °C, 3.3V |
±5 |
nA |
||
Current leakage (Current + to -) |
25 °C, 5V |
±7 |
nA |
||
Current leakage (Current + to -) |
25 °C, 15V |
±28 |
±5000 |
nA |
|
Max fuse engage |
5 A |
0.293 |
s |
||
Max fuse engage |
10 A |
0.050 |
s |
General purpose inputs and outputs#
Test condition |
Min |
Typical |
Max |
Unit |
|
|---|---|---|---|---|---|
Internal 3.3V reference |
3.135 |
3.3 |
3.465 |
V |
|
External reference |
Vref < (VUSB - 0.5 V) |
Vref - 0.1 |
Vref |
Vref + 0.1 |
V |
Input leakage per pin |
25 °C |
±10 |
nA |
||
Input leakage per pin |
±100 |
nA |
|||
Output voltage high |
-100 µA |
Vcco - 0.1 |
V |
||
Output impedance |
10 mA |
45 |
50 |
80 |
Ω |
GPI sample rate |
2 |
MSPS |
|||
Continuous output current |
Per pin |
±50 |
mA |
||
Total power dissipation |
Over output impedance |
0.25 |
W |
The Joulescope uses the TI SN74LVC2T45 for both the GPIs and GPOs. See the SN74LVC2T45 datasheet for more information. Both the GPIs and GPOs are protected with the TI TPD4E1U06 for ESD protection.