2. Specifications#

This section only gives the specification changes relative to the Joulescope JS220 User’s Guide. All other specifications remain the same as the JS220. The JS220+ datasheet lists the complete specifications.

2.1. Current measurement specifications#

This table replaces the JS220 current measurement specifications.

Range

Accuracy

Resolution

Impedance

Guaranteed

10 A (> 2 A)1

±0.25% ±1.5 mA

350 µA

0.01 Ω

±9.7 A

10 A (≤ 2 A)

±0.1% ±300 µA

75 µA

0.01 Ω

±2.05 A

180 mA

±0.1% ±30 µA

6.5 µA

0.11 Ω

±185 mA

18 mA

±0.1% ±3.0 µA

650 nA

1.11 Ω

±18.5 mA

1.8 mA

±0.1% ±300 nA

65 nA

11.1 Ω

±1.85 mA

180 µA

±0.1% ±30 nA

6.5 nA

111 Ω

±185 µA

18 µA

±0.1% ±3 nA2

0.65 nA

1111 Ω

±18.5 µA

  1. Max 3 A sustained. See the recommended operating conditions for higher current durations.

  2. ±1.0 nA for ±5 °C over 24 hours with trim offset calibration.

2.2. Voltage measurement specifications#

This table replaces the JS220 voltage measurement specifications.

Range

Accuracy

Resolution

Guaranteed

15 V

±0.1% ±1 mV

500 µV

±14.5 V

2 V

±0.1% ±150 µV

75 µV

±2.2 V

2.3. Parasitic effects specifications#

This table replaces the JS220 parasitic effects specifications.

Test condition

Min

Typical

Max

Unit

Current leakage (Voltage + to -)

25 °C, ±3.3V

200

pA

Current leakage (Voltage + to -)

25 °C, ±5V

200

pA

Current leakage (Voltage + to -)

25 °C, ±14V

300

pA

Shunt resistor voltage, max

-2 to 2 A, static

20

21

mV

Burden voltage (Current + to -)

1 A, banana jacks

0

25

35

mV

Burden voltage (Current + to -)

Shunt resistor, < ±2 A

±20

mV

Shunt switching time

Over-range, < 200 ns rise time

0

1.0

1.5

µs

Measurement gap following a current range switch

0

µs

Input bias current Voltage+ & -

25 °C

±5

±20

pA

Input bias current, Current + & -

25 °C, 18 µA current range

±60

±600

pA

Resolution is given for 10 kHz measurement bandwidth (20 ksps).

The JS220+ includes a safety banana jack front panel rather than binding posts.